5-10 Setting the Measurement Conditions for the Program
1
Click on the [Area settings] tab in the
program setting area.
The [Area settings] window appears.
2
Set which area of the target to measure.
Area settings
Select the measurement method from the pull-down menu.
Measurement
method
Type of measurement
DIA Set this method when measuring the outer diameter of a round bar or a transparent body. This method
measures the target from the first falling edge (an edge where a bright portion becomes dark) to the last
rising edge (an edge where a dark portion becomes bright).
T- E D G E Set this method when measuring the gap between rollers or when setting a position. T-EDGE measures the
width of the first bright portion on the top of the target.
B-EDG
E Set this
method when measuring the gap between rollers or when setting a position. B-EDGE measures the
width of the last bright portion on the top of the target.
SEG Set this method when measuring the inner diameter of a disk or when measuring pitch.
This method measures the width between two specified edges: the start edge where measurement is
started and the end edge where measurement is finished.
Parameter range: 0, ±1 to ±127, E or P
Copies the calibration settings from area 1
or area 2 and pastes them into area 1 or
area 2.
1
Set the detection threshold value
when measuring a target that has high
transparency. Enter a value into the
text box.
Parameter range: 10 to 99%
Set two areas
(area 1 or area
2) on the
measurement
target.
Select [ON] or [OFF] from the pull-down
menu. When the parameter is set to [ON],
set the total number of edges for
measurement. When the value has a
different number of edges, the
measurement value becomes invalid. This
prevents the output of measurement values
that are incorrect due to dust or oil on the
measurement head.
Parameter range: 2 to 127
Select the head (head 1 or head 2) to use
for measurement.