6-11
Chapter 6 I/O Terminals
6
Timing Chart
This section provides information on the timing charts of BCD output in a variety of meas-
uring modes.
Normal, Auto Peak Hold, Auto Bottom Hold, and Auto Peak-to-peak
Hold
ON
OFF
TIMING input
ON
OFF
RESET/ZERO input
ON
OFF
P1 input to P4 input
ON
OFF
Comparator output
ON
OFF
WAITING output
ON
OFF
ON
OFF
STROBE output
BCD output
Refresh cycle
STROBE time
0±0.1ms
Min Min2.1ms
Min2.1ms
Max2.5ms
RESET
time
Max2.5ms Max2.5ms
2.1ms
Min
2.1ms
Min 6.0ms
Max 8.0ms
Min
6.0ms
RESET
time
The refresh cycle and reset time vary with the setting of the averaging number.
Refer to Reset Time and Refresh Cycle on page 6-13 for detail.
The strobe output time can be changed. If the time set exceeds the refresh cycle time,
the strobe output will be always ON.
Refer to Strobe Output Time on page 4-48 for detail.
Peak Hold, Bottom Hold, Peak-to-peak Hold, Average Hold, and
Sample Hold 1
ON
OFF
TIMING input
ON
OFF
RESET/ZERO input
ON
OFF
P1 input to P4 input
ON
OFF
Comparator output
ON
OFF
WAITING output
ON
OFF
ON
OFF
STROBE output
BCD output
STROBE time
0±0.1ms
Min2.1msMin
2.1ms
Min2.1ms
Min2.1ms
Max2.5ms RESET
time
RESET
time
Sampling
period
Sampling
period
Max2.5msMax2.5ms
Min 6.0ms
Min
6.0ms
Max 8.0ms
The strobe output time can be changed.
Refer to Strobe Output Time on page 4-48 for detail.
Reset time varies with the setting of the averaging number.
Refer to Reset Time and Refresh Cycle on page 6-13 for detail.
When RESET input turns ON, the internal measured value will be determined with an
elapse of reset time.