8 AVE Key
(Selector key for number
of measurements for
averaging and averaging
method)
Sets the desired number of measurements and averaging method.
1. Use AVE to display the current number and method.
2. To change the number of measurements, use UP/DOWN to move the
cursor up. Then, use SET to specify the number. Each time you press
SET, the number will increment in the following order: "1, 2, 4, 8 ••••• 512,
and 1024". Note also that the number of measurements is set automati-
cally to the currently displayed number. Therefore, you do not need to
press any other key to finalize this number.
3. To change the measurement method, use UP/DOWN to move the cursor
down. Then, use SET to change the averaging method. Each time you
press SET, SIMPLE and MOVE display alternately.
See the timing diagrams below for the difference between these two
methods.
CHAPTER 4 Functions and Controls
AVE
1024
MOV E
AVE
1024
SIMPLE
* MOVE is valid only when the number of measurements is set to a number
between 16 and 1024.
Averaging Methods
1, 2, 3, • • • • • • • 30, 31, 32, 33, 34, 35, • • • • • • 62, 63, 64,65, 66, 67, • • • 94, 95, 96, 97, 98, 99, • • • 126, 127, 128, 129, 130, 131, • • •
1, 2, • • 7, 8, 9, 10, • • 15, 16,17, 18, • • 23, 24, 25,26, • • 31, 32, 33, 34, • 39, 40, 41, 42, • 47, 48, 49, 50, • 55, 56, 57, 58, • •
1
23 4
1
2
3
4
1
2 34
1
23
4
Number of samplings
SIMPLE averaging
(Number: 32)
Output cycle of measured value
2.5 ms x number of measurements
Number of samplings
MOVE averaging
(Number: 32)
2.5 ms x number of measurements
2.5 ms x number of measurements
20 ms
Out put cycle of measured value