CHAPTER 4 Functions and Controls
Procedure Check first that segment pair, (m,n), has been selected.
1. Use FNC to display "X SEG CHECK OFF".
2. Use SET to display "ON" on the right of "SEG CHECK" to activate the
segment check mode.
* When the currently selected segment pair is DIA, EDGE1, or EDGE2, "Y
SEG CHECK" will be displayed.
3. Enter the correct number of edges.
4. Press ENT to register.
5. Press FNC to display "Y SEG CHECK OFF".
6. Repeat steps 3 and 4.
Example of Operation
SCEGX HECK OFF
4
SCEGX HECK ON
4
SCEGY HECK OFF
4
SCEGY HECK ON
6
[
]
SCEGX HECK ON
4
SCEGX HECK ON
6
[]
SCEGX HECK ON
6
SCEGY HECK ON
4
1
2
3
4
5
6
Segment selectors
X=(2. 3)
Y=(4. 5)
ReceiverTransmitter
SCEGY HECK ON
6
You are to measure the outer diameter of two round bars at the same time
using one scanning head. You need to use the segment check mode to
prevent incorrect measurements. In this case, the number of edges is 6.
Step Key operation Display
1 Use FNC to display "X SEG
CHECK OFF".
2 Press SET.
3 Enter "6".
4 Press ENT to register.
5 Press FNC.
6 Press SET.
7 Press "6".
8 Press ENT to register.
SCEGX HECK OFF
4
ls3100_30_48.p65 08.10.1, 10:55 AM36