4.3 Command Communication
53
4.3.4 Response Error Code
When an incorrect command is sent to set parameters, the BL-180 sends back
data indicating the cause of the error (error code). For the commands correspond-
ing to the error codes, see the error code column given in the table on the follow-
ing pages.
4.3.5 Description of Parameter Setting Commands
Saving/Initializing Settings/Quitting Setting Mode
Error code Cause of error
00 Undened command.
01 Command format is incorrect.
02 Nothing corresponds to the number in the command.
03 m value (codes 1 to 4) is other than 0 to 3.
04
Bar code type setting command was not sent rst.
Sending No. of readable digits setting command is invalid
for JAN code.
05 The number in the command is too long.
06 hhh... data is too short.
07 n value is not 0 or 1.
08 n value is exceeding the setting range.
09 nnn or nn value is exceeding the setting range.
10 hhh... is not specied in HEX (hexadecimal) code.
11 hhh...=FF cannot be set.
12
hhh... or aaa... contains more than the specied number
of characters.
13 Characters of aaa... are invalid.
14
Data in the EEPROM may be damaged. Perform initial
setup.
15
Error in the area storing initial settings.
Settings are automatically initialized.
17 \ is not followed by !, ? or \ in preset data. (See page 70.)
18 Two !s exist in preset data. (See page 20.)
99 The BL-180 may malfunction. Contact KEYENCE.
Function Command being sent
Respons
e
Description
Error
Code
Saves settings in the EEP-
ROM.
SAVE OK See page 51. -
Initializes settings. DFLT OK Returns to the default-settings, and saves the set-
tings in the EEP-ROM.
00,05
14,15
Quits the setting mode. SEND OK See page 51. -