67
Error Messages
(For Models LS-3060 and
LS-3033)
Display Description
In the POSITION CHECK mode, the target position indicator LEDs show
whether the transmitter and receiver (scanning head) are aligned in terms of
the optical axis.
The transmitter and receiver are aligned .
"P-Err" appears, and LED 3 or 5 lights.
"P-Err" appears, and LED 1 or 7 lights.
When the transmitter is tilted with the laser emitting
surface facing upward as shown above, LED 1 lights.
When the transmitter is inclined with the surface facing downward,
LED 7 lights.
"P-Err" appears, and all the LEDs light.
When the transmitter and receiver are positioned as
shown above, all the LEDs light.
CHAPTER 5 Mounting Scanning Head
When the transmitter is positioned higher than the
receiver as shown above, LED 3 lights.
When the transmitter is positioned lower than the receiver, LED 5
lights.
"GOOD" appears, and target position indicator LED
4 lights.
OFFSET LOCK POSITION
mm
inch
123 4 567
OFFSET LOCK POSITION
inch
2314567
mm
OFFSET LOCK POSITION
inch
2314567
mm
OFFSET LOCK POSITION
inch
2314567
mm